Find out how to change a UV lamp and its quartz on UVRER's new DFI range.
-
1.Take out the lamp
- Unplug your device,
- Unscrew the lamp,
- Carefully take it out of the device,
- Remove the centering ring.
-
2.Remove the quartz
- Remove the cable gland,
- The flat gasket,
- The O-ring,
- Then the quartz sheath, delicately.
-
3.Put the new quartz
Repeat the operation in reverse:
- Insert the quartz sheath by holding it with two fingers inside,
- Put the O-ring,
- The flat gasket,
- The stuffing box.
-
4.Put in the new lamp
- Tighten the centering ring at the end of the device,
- Insert the lamp and screw it.